Sharp 4K Arthroscope 30°AR-3350 4031 Repair Case

1. Fault content: As identified by our experienced endoscope repair specialist, the endoscope exhibited multiple critical failures, including no image output, a broken objective lens, water leakage, a damaged rod lens, and a broken connection between the tube and the main body. These issues significantly affected both imaging performance and operational safety.


2. Repair plan: As a professional endoscopy equipment supplier and repair service provider, we implemented a comprehensive repair plan that included replacing the objective lens, rod lens, tube, and light guide (complete overhaul), followed by precise image calibration. After repairs, the endoscope underwent rigorous testing, including anti-vibration and impact resistance testing, sealing performance verification, and hot and cold shock testing, to ensure reliable performance and compliance with clinical use standards.