Tel: 4008-266-720
E-mail: info@sihanmed.com
– Smooth cutting through different tissue types, offering both Pure Cut for scalpel-like cuts and Blend for hemostasis while cutting
– Coagulation for both dissection and hemostasis. Choose either Standard Coag for dissection with hemostasis or Spray Coag for broader hemostasis
– Precise bipolar dessication provides efficient dessication with minimal lateral thermal tissue damage. Choose between Micro for fine and precise coagulation, and Macro for energy intensive coagulation
– Automatic programming restores the System 2450 to the last settings used
– One handswitch/footswitch combination jack and a separate handswitch jack allows for connection of multiple accessories
– Serial interface port enables advanced communications between the System 2450 and compatible robotic surgical systems or OR integration systems
Output Characteristics (Maximum Output)
Gross F | Maximum output | Fixed Load(Ω) | Representative of 7L and F79 I (50W eye) | Maximum positive displacement (C-7) | Transmission Wave Number(KHz) |
Pure stomach cutting | 300W | 500Ω | 1.4~2.0 | 1100V | 390-410 |
Mixed Purging | 200W | 500Ω | 2.0-2.6 | 1200V | 390-410 |
Coagulation | 120W | 500Ω | 5.6--6.6 | 3200V | 485-505 |
Coagulation | 80W | 500Ω | 7.5-9.5 | 4400V | 485-505 |
Coagulation | 70W | 50Ω | 1.4-2.0 | 150V | 390-410 |
Coagulation | 70W | 300Ω | 1.4--2.0 | 560V | 390-410 |
Input Force Characteristics (Output Force Maximum Setting Load 500Ω)
Power supply | 100V AC±10% 50/60Hz |
Current | 8.4A (CUT max.) |
Power consumption | 840VA |
High frequency leakage current
Mode | Scalpel tip electrode - ground |
Monopolar incision and coagulation | <100mA |
General
Form of protection against electric shock | Class I equipment |
Degree of protection against electric shock | CF type |
Drip-proof | Drip-proof type |
Internal design | Solid state |
Output method | Floating type |
Cooling method | Natural heat dissipation type (without fan) |
Dimensions | |
Width | 37cm |
Depth | 48cm *including handle |
Height | 18cm |
Weight | 14.5kg |
Mains power overcurrent protection | Fuse 10.0A*Two type T fuses are used, high interrupting |
Alarm volume | 65db (min.) |
Output volume | 45db (min.) |
Operating conditions | 10°C to 30°C, 95% relative humidity |
Duty cycle | 15 sec ON / 30 sec OFF |