Repair Case of Wolf Camera 85525902

Focusing lens: 85261504

Focusing lens: 5000380997     

Second repair

Appearance inspection: The appearance of the camera shows signs of normal use and wear. The cable sheath at the camera end is seriously damaged, and the overall cable appearance shows aging, as identified by a professional endoscope repair specialist during the detailed inspection.

Fault detection: Tested on the machine, the color of the picture has a border. Connect a mirror to observe the palm of the hand, there is color distortion. Replace the spare CCD for testing, and the picture is normal.

Fault cause: The CCD is damaged, resulting in the color of the picture having a border.

Repair plan: Conduct professional surgical camera repair by replacing the CCD and installing a new cable to restore full image quality and reliable performance.