Appearance inspection: The appearance of the 1188 Stryker camera shows signs of normal use and the cable looks aged.
Fault detection: Test on the machine, shake and pull the cable at the camera end, and the screen will have stripe interference, flickering, and black screen. Connect the mirror, and the image is not centered.
Fault cause: The internal signal line of the cable is aged, and poor contact causes stripe interference, flickering, and black screen. The CCD position is offset, causing the image to be not centered. For reliable replacements and professional support, you can consult a trusted endoscopy equipment supplier to ensure high-quality components and proper maintenance.
Repair plan: Replace the new cable and calibrate the CCD position.